How do I perform a manual transmission check?

How do I perform a manual transmission check?

Open the SoloVPE software.

Click on  “Manual” located in the sidebar menu.

Disconnect the delivery fiber from the SoloVPE translation stage by unscrewing it counter-clockwise and then completely remove it by pulling it out from the top of the strain relief located above the SoloVPE.

Obtain the transmission tool from your SoloVPE accessory kit.

Connect the delivery fiber to the transmission tool. Make sure to align the key before you tighten the screw.

Place the connected transmission tool directly on the SoloVPE Detector window.

From the manual screen locate the “Y Mode” option and Switch it to ‘%T’.

Press ‘Quick Read’.

The transmission results will be printed in the report window.